International Journal of Medical Biochemistry

[Int J Med Biochem ]
Int J Med Biochem . 2023; 6(3): 165-172 | DOI: 10.14744/ijmb.2023.19042  

The use of new generation small-volume blood collection tubes for complete blood count

Ahmet Erkin Bozdemir1, Fatma Demet Arslan2, Banu Isbılen Basok1, Sukran Copur3, Nisel Ozkalay Yilmaz3, Harun Akar4
1Department of Medical Biochemistry, University of Health Sciences, Tepecik Training and Research Hospital, Izmir, Turkiye
2Department of Medical Biochemistry, Bakircay University, Cigli Training and Research Hospital, Izmir, Turkiye
3Department of Medical Microbiology, University of Health Sciences, Tepecik Training and Research Hospital, Izmir, Turkiye
4Department of Internal Medicine, University of Health Sciences, Tepecik Training and Research Hospital, Izmir, Turkiye

INTRODUCTION: Newborns and especially patients with malignancy develop frequently iatrogenic anemia based on phlebotomy. The use of blood collection tubes with small-volume may reduce the need for blood transfusion and the associated risks due to frequent phlebotomy. Therefore, we evaluated the reliability and accuracy of the complete blood count (CBC) using new generation small-volume blood collection tubes (SV-BCT) instead of large-volume blood collection tubes (LV-BCT).
METHODS: Venous blood samples were taken from 40 adult in-patients and collected to SV-BCT/LV-BCT pairs of three different brand (Microtainer®MAP-0.5 ml/Vacutainer®-2.0 ml; Becton, Dickinson and Company, USA) ([Microvette®-0.5 mL/S-Monovette®-2.6 ml; Sarstedt Ag and Co. KG, Germany]) ([MiniCollect®Complete-0.5 ml/Vacuette®-2.0 ml; Greiner Bio-One GmbH, Austria]). All tubes contained K2EDTA except Microvette®. Sixteen parameters of CBC were analyzed using a DxH 800 (Beckman Coulter Inc., USA). CBC results in tube pairs were compared in terms of statistical and clinical (bias%) significance.
RESULTS: There were statistically significant differences between the results of SV-BCT and LV-BCT pairs of the same brands for some parameters. However, bias% between tubes for 16 parameters was within the desirable limits, the differences were not clinically significant.
DISCUSSION AND CONCLUSION: Health personnel and phlebotomists can safely prefer SV-BCT which is a new generation and technically useful for CBC, especially in patients requiring frequent phlebotomy. Thus, the volume of blood sampling may be reduced to prevent iatrogenic blood loss.

Keywords: Blood collection tube, complete blood count, Iatrogenic anemia, phlebotomy


Ahmet Erkin Bozdemir, Fatma Demet Arslan, Banu Isbılen Basok, Sukran Copur, Nisel Ozkalay Yilmaz, Harun Akar. The use of new generation small-volume blood collection tubes for complete blood count. Int J Med Biochem . 2023; 6(3): 165-172

Sorumlu Yazar: Ahmet Erkin Bozdemir, Türkiye


ARAÇLAR
Tam Metin PDF
Yazdır
Alıntıyı İndir
RIS
EndNote
BibTex
Medlars
Procite
Reference Manager
E-Postala
Paylaş
Yazara e-posta gönder

Benzer makaleler
PubMed
Google Scholar